%Linearity per AIAG (MSA Type 4)

AIAG KPI for the linearity of a measurement system across its entire range.

Statusvalidated
Version1.0.0
Minitab equivalentStat > Quality Tools > Gage Study > Gage Linearity and Bias Study → %Linearity

Description

From the linearity regression (slope a) and the range of reference values, %Lin = 100 · |a| · Range / PV is computed. PV is the tolerance T = USL − LSL (or 6·σ_P). Additionally %Bias_max = 100 · max|b_i| / PV is reported. Traffic light: green < 5 %, yellow 5–10 %, red > 10 % or a significant slope/intercept.

Formulas

%Linearität
AIAG definition: absolute slope times reference range, normalized by the process variation
%Bias-Peak
Largest single-point bias across all references, normalized by PV
Range
Range of the reference values used in the study
PV
Reference figure for the normalization — tolerance or 6·σ_P depending on pvMode

Assumptions

  • The regression assumptions (bias normality, homoscedasticity) are satisfied
  • Reference values cover the relevant measurement range between LSL and USL

Limitations

  • For very small ranges %Lin is underestimated — the test loses discriminative power
  • AIAG thresholds 5/10 % are conventions — customers may require stricter limits

References

  • AIAG — Measurement Systems Analysis Reference Manual, 4th Edition (MSA-4), Kapitel III-B: Linearity
  • Dietrich, E. & Schulze, A. — Statistische Verfahren zur Maschinen- und Prozessqualifikation, 7. Auflage, Hanser