MSA Type 1 (Cg/Cgk)
Measurement System Analysis Type 1 — Repeatability (Cg) and Bias (Cgk)
Overview
The MSA Type 1 study checks a measurement system for bias (position) and repeatability (spread) — on one single reference part, with one operator, in a short time frame. It is the first step of any measurement-system analysis: if it fails, no deeper study (Type 2 / Gage R&R) is worthwhile (AIAG 2010).
Reference part: A reference part or master with a known, traceable reference value. Without this value, position cannot be judged — the bias part of the study would be blind.
Repeat measurements: Typically 25–50 measurements of the same reference with the same system under conditions as constant as possible. Only then does pure repeat variation become visible.
Bias: Systematic deviation: mean of repeat measurements − reference value. A bias ≠ 0 indicates a calibration offset or systematic shift and breaks the trueness of the system.
Cg: Cg measures repeatability relative to tolerance. Typical formula: Cg = 0.2·T / (6·s), where T is the tolerance width and s the standard deviation of repeats. Accepted when Cg ≥ 1.33 (Verband der Automobilindustrie 2021).
Cgk: Cgk extends Cg to include bias. Cgk ≥ 1.33 is accepted (Verband der Automobilindustrie 2021). High Cg, low Cgk → the system is precise but shifted (calibration issue).
k₂ (spread width): Default 3 = 6·s as in Bosch Heft 10 and Minitab — the standard value for this study. 4 (8·s) and 6 (12·s) are stricter alternatives with a wider spread. Studies already saved keep their previous setting.
%Var: %Var(repeatability) = 100·k₁/Cg and %Var(repeatability and bias) = 100·k₁/Cgk express the Cg/Cgk acceptance limit of 1.33 as a share of tolerance consumed by spread: Cg = 1.33 corresponds to %Var(repeatability) ≈ 15%.
Bias t-test: t = bias / (s / √n), degrees of freedom df = n − 1. p < 0.05 means the bias is statistically significant. The test is informational — the verdict (capable / not capable) is based on Cg and Cgk alone.
One-sided tolerance: with only an upper or a lower specification limit given, Cg and Cgk cannot be computed (as in Minitab) — the run chart, statistics, and bias t-test are still shown.
Type 1 does not replace Gage R&R, it is the pre-test. Failing Type 1 means running a Gage R&R afterwards wastes resources.
Approach
- Provide a reference part with a traceable value — ideally near mid-spec.
- Use the system under production conditions, not lab conditions.
- Take 25–50 repeat measurements on one day without re-setup.
- Compute mean, standard deviation, and bias.
- Compute Cg and Cgk and compare to the acceptance limit (usually 1.33).
- Document the result; if it fails, investigate causes before running Gage R&R.
Pitfalls
No traceable reference value: A reference part without a known value disables bias assessment — the study then measures only repeatability. Without a reference, the result is incomplete.
Optimistic lab conditions: Testing a gage in air-conditioned calm produces nice Cgk — irreproducible in real production. The test must mirror actual use.
Too few measurements: With 10 repeats, s and bias are uncertain, Cg/Cgk unstable. 25–50 repeats are standard; below that it becomes unreliable.
Measurements "optimized": Freely dropping outliers produces nice numbers and poor reality. Record all measurements; remove outliers only with a documented cause.
Type 1 passed = system OK: Type 1 only checks bias and repeatability on one part by one operator. Variation from operators or part-to-part range shows up only in Type 2 / Gage R&R.
Acceptance threshold too lax: Cg/Cgk < 1.33 is not internationally accepted. Lowering the bar internally makes the study worthless the moment a customer audit appears.
References
- AIAG (2010)Measurement Systems Analysis (MSA) Reference Manual4th ed., Automotive Industry Action Group, Southfield, MI, ISBN 978-1-60534-211-5Basis for bias, repeatability and the acceptance practice for capability indices.
- ISO (2021)ISO 22514-7: Capability of measurement processesInternational Organization for Standardizationhttps://www.iso.org/standard/78307.htmlNormative framework for the capability of measurement processes.
- Melzer, Almut (2019)Six Sigma – kompakt und praxisnah: Prozessverbesserung effizient und erfolgreich implementieren2nd ed., Springer Gabler, Wiesbaden, ISBN 978-3-658-23754-7doi:10.1007/978-3-658-23755-4Cross-check of the type 1 study: 6·s default, bias t-test, %Var(repeatability) and one-sided tolerance.
- Robert Bosch GmbH (2019)Qualitätsmanagement in der Bosch-Gruppe, Technische Statistik, Heft 10: Fähigkeit von Mess- und PrüfprozessenRobert Bosch GmbH, StuttgartDefault k₂ = 3 (6·s) for Cg and Cgk, as in Minitab.
- Verband der Automobilindustrie (2021)VDA Band 5 — Mess- und Prüfprozesseignung3rd ed., VDA QMC, BerlinDefines Cg and Cgk via 0.2·T and 6·s, and the 1.33 acceptance limit.
Examples
This module ships with the following example datasets — load any of them in the app with a single click.
Available in the following cycles
- DMAIC: Measure
- DMADV: Measure